Analytical Sample Preparation

Aliquots of dissolved proteins (plasma, CSF, cell lysate) were diluted with sodium bicarbonate in concentration 1ug/ul. Protein solution was reduced with 5 mM DTT for 60 min at 60°C and alkylated with 15 mM iodoacetamide for 30 min in the dark. Trypsin Gold (Promega, Madison, WI) digestion (2.5 ng/μl) was carried out at 37°C in Barocycler NEP2320 (Pressure BioSciences, South Easton, MA) for 1 hour.

Chromatography

Digested proteins were separated using a 90 minute ACN gradient on a 250 mm x 75 μm C18 pepmap column at a flow rate of 0.3 μL/min. In brief, peptide and glycopeptide separation was achieved by a 5 min trapping/washing step using 99% solvent A (2% acetonitrile, 0.1% formic acid) at 10 μL/min followed by a 90 min acetonitrile gradient at a flow rate of 0.3 μL/min: 0-3 min 2% B (0.1% formic acid in ACN), 3-5 min 2-10% B; 5-60 min 10-45% B; 60-65 min 45-98% B; 65-70 min 98% B, 70-90 min equilibration by 2% B.

Mass Spectrometry

Glycopeptides were analyzed by DIA method using Orbitrap Lumos Fusion mass spectrometer. The electrospray ionization voltage was 3 kV, the capillary temperature 275°C. MS1 scans were performed over m/z 400–1800 with the wide quadrupole isolation 20Da, RF Lens at 40%, intensity Data-independent HCD tandem mass spectra were collected with a resolution of 30,000 in Orbitrap with range 150-1800 and normalized collision energy 20%.

Acquisitions

SampleSpectrum FileLC Cal. SlopeLC Cal. Int.Trans. Groups
Human SerumMS_47_UO1_SERUM_DIA-15_prof2.459-29.325176